This divider is two 10 MΩ resistors across 9 V. Huge values, but a divider only cares about the ratio.

What voltage should its midpoint be at?

Answer in V.

Open the multimeter, put red on the midpoint (column 7) and black on the − rail, and press Run. What does it read?

Answer in V.

A voltmeter measures by letting a tiny current through itself. Inside, between its probes, is a resistance: 10 MΩ for a typical digital meter. Usually that's so big it might as well not be there.

Why did it read 3.00 V here?

  • Its 10 MΩ is in parallel with the bottom 10 MΩ, making 5 MΩ, so the divider becomes 10 MΩ over 5 MΩ: a third of 9 V
  • The meter is faulty
  • 10 MΩ resistors are inaccurate

Here is the same divider made from 10 kΩ resistors, with the meter already on its midpoint. Press Run. What does it read now?

Answer in V.

Seen from the meter, any circuit is a Thévenin equivalent: Vth behind Rth. The meter's 10 MΩ and Rth form a divider, so the meter shows

Vread=Vth×10 MΩRth+10 MΩ

A sensor's output has a resistance of 2.2 MΩ. A meter across it reads 2.46 V. What is the sensor's true output voltage, with no meter connected?

Answer in V.

Roughly when can you trust a 10 MΩ voltmeter to within 1%?

  • When the circuit's resistance at that point is under about 100 kΩ: a hundredth of the meter's
  • Always: 10 MΩ is enormous
  • Only in circuits with no resistors over 1 kΩ

Meters are not invisible:

  • a voltmeter is a big resistor, typically 10 MΩ, across what it measures;
  • on a high-resistance circuit it loads the point it measures and reads low;
  • the error is about Rth ÷ 10 MΩ: negligible below 100 kΩ;
  • knowing Rth, correct a reading: Vth=Vread(Rth+Rm)/Rm.